Authors
Ian K Webb, Sandilya VB Garimella, Aleksey V Tolmachev, Tsung-Chi Chen, Xinyu Zhang, Randolph V Norheim, Spencer A Prost, Brian LaMarche, Gordon A Anderson, Yehia M Ibrahim, Richard D Smith
Publication date
2014/9/16
Journal
Analytical chemistry
Volume
86
Issue
18
Pages
9169-9176
Publisher
American Chemical Society
Description
We report on the performance of structures for lossless ion manipulation (SLIM) as a means for transmitting ions and performing ion mobility separations (IMS). Ions were successfully transferred from an electrospray ionization (ESI) source to the TOF MS analyzer by means of a linear SLIM, demonstrating lossless ion transmission and an alternative arrangement including a 90° turn. First, the linear geometry was optimized for radial confinement by tuning RF on the central “rung” electrodes and potentials on the DC-only guard electrodes. Selecting an appropriate DC guard bias (2–6 V) and RF amplitude (≥160 Vp-p at 750 kHz) resulted in the greatest ion intensities. Close to ideal IMS resolving power was maintained over a significant range of applied voltages. Second, the 90° turn was optimized for radial confinement by tuning RF on the rung electrodes and DC on the guard electrodes. However, both resolving …
Total citations
2014201520162017201820192020202120222023202437141491791111149