Authors
Robert W Cox, Andrzej Jesmanowicz, James S Hyde
Publication date
1995/2
Journal
Magnetic resonance in medicine
Volume
33
Issue
2
Pages
230-236
Publisher
Wiley Subscription Services, Inc., A Wiley Company
Description
A recursive algorithm suitable for functional magnetic resonance imaging (FMRI) calculations is presented. The correlation coefficient of a time course of images with a reference time series, with the mean and any linear trend projected out, may be computed with 22 operations per voxel, per image; the storage overhead is four numbers per voxel. A statistical model for the FMRI signal is presented, and thresholds for the correlation coefficient are derived from it. Selected images from the first real‐time functional neuroimaging experiment (at 3 Tesla) are presented. Using a 50‐MHz workstation equipped with a 14‐bit analog‐to‐digital converter, each echo planar image was acquired, reconstructed, correlated, thresh‐olded, and displayed in pseudocolor (highlighting active regions in the brain) within 500 ms of the RF pulse.
Total citations
19951996199719981999200020012002200320042005200620072008200920102011201220132014201520162017201820192020202120222023202427820131922151117137111361015222029122227221114141565
Scholar articles
RW Cox, A Jesmanowicz, JS Hyde - Magnetic resonance in medicine, 1995