Authors
Robert W Spitz, Yujiro Yamada, Vickie Wong, Ryo Kataoka, William B Hammert, Jun Seob Song, Anna Kang, Aldo Seffrin, Jeremy P Loenneke
Publication date
2024/2/26
Journal
Physiological Measurement
Volume
45
Issue
2
Pages
02NT01
Publisher
IOP Publishing
Description
Blood flow restriction pressures are set relative to the lowest pressure needed to occlude blood flow with that specific cuff. Due to pressure limitations of some devices, it is often not possible to occlude blood flow in all subjects and apply a known relative pressure in the lower body with a 5 cm wide cuff.
Objective
To use a device capable of generating high pressures (up to 907 mmHg) to create and validate an estimation equation for the 5 cm cuff in the lower body using a 12 cm cuff.
Approach
170 participants had their arterial occlusion pressure (AOP) with a 5 cm and 12 cm cuff and their thigh circumference measured in their right leg. The sample was randomly allocated to a prediction group (66%) and validation group (33%). Thigh circumference and 12 cm AOP were used as predictors. A Bland–Altman plot was constructed to assess agreement between measured and predicted values.
Main results
The mean …