Authors
Bharathi Avula, Yan-Hong Wang, Troy J Smillie, Wilfred Mabusela, Leszek Vincent, Frans Weitz, Ikhlas A Khan
Publication date
2009/3/1
Journal
Journal of AOAC International
Volume
92
Issue
2
Pages
633-644
Publisher
Oxford University Press
Description
A simple and specific analytical method for the quantitative determination of flavonoids from the aerial parts of the Artemisia afra plant samples was developed. By column high-performance liquid chromatography (HPLC) with UV absorption and mass spectrometry (MS) detection, separation was achieved on a reversed-phase octadecylsilyl (C18) column with water, methanol, and acetonitrile, all containing 0.1 acetic acid, as the mobile phase. These methods were used to analyze various species of Artemisia plant samples. The wavelength used for quantification of flavonoids with the diode array detector was 335 nm. The limits of detection (LOD) by HPLC/MS were found to be 7.5, 7.5, 10, 2.0, and 2.0 ng/mL; and by LC-UV the LODs were 500, 500, 500, 300, and 300 ng/mL for apigenin, chrysoeriol, tamarixetin, acacetin, and genkwanin, respectively. The HPLC/MS method was found to be 50150 times more …
Total citations
2010201120122013201420152016201720182019202020212022202322223414454324