During hyperthermia the presence of a large vessel entering the heated volume and carrying blood at the systematic temperature can be an important source of temperature non-uniformity and possible underdosage. The minimal tumour temperature near a large vessel is determined by the vessel wall temperature: a number of factors influencing the vessel wall temperature are considered-effective tissue conductivity, flow type, vessel size, entrance effects and counter-current flow. In some specific cases, especially when tissue perfusion is high, the vessel wall temperature may reach therapeutic levels when the mean blood temperature is still low. In general, well perfused tumours have a better chance of being heated uniformly. Regional heating improves temperature uniformity by reducing entrance and equilibration effects as blood is heated before entering the tumour. Raising the core temperature also reduces …