Authors
Y Yang, SF Zhang, MA Kingston, G Jones, G Wright, SA Spencer
Publication date
2000/8/1
Journal
Biosensors and bioelectronics
Volume
15
Issue
5-6
Pages
221-227
Publisher
Elsevier
Description
A new biocompatible copolymer has been synthesised and used in an electrochemical enzyme-based glucose sensor. The copolymer incorporates three segments including a monomer with an electrically neutral phosphorylcholine head group that is able to reject protein adsorption and two segments that increase the affinity to polyurethane substrate. Peel and solution circulation tests showed that this material has high attachment to polyurethane. With the new copolymer as the outermost layer and the polyurethane as the diffusion-limiting membrane, the sensor showed extended linearity up to 50 mM glucose and stable output in bovine serum for 70 h. During in vivo tests, the sensor exhibited a steady current signal and a rapid transient response when the glucose concentration was raised. These results imply that the haemocompatibility of the glucose sensor coated with the new copolymer has been improved …
Total citations
200120022003200420052006200720082009201020112012201320142015201620172018201920202021202235142636233444351232
Scholar articles
Y Yang, SF Zhang, MA Kingston, G Jones, G Wright… - Biosensors and bioelectronics, 2000