Authors
Xiaokun Liang, Zhicheng Zhang, Tianye Niu, Shaode Yu, Shibin Wu, Zhicheng Li, Huailing Zhang, Yaoqin Xie
Publication date
2017/6/6
Journal
Physics in Medicine & Biology
Volume
62
Issue
13
Pages
5276
Publisher
IOP Publishing
Description
Ring artifacts in cone beam computed tomography (CBCT) images are caused by pixel gain variations using flat-panel detectors, and may lead to structured non-uniformities and deterioration of image quality. The purpose of this study is to propose a method of general ring artifact removal in CBCT images. This method is based on the polar coordinate system, where the ring artifacts manifest as stripe artifacts. Using relative total variation, the CBCT images are first smoothed to generate template images with fewer image details and ring artifacts. By subtracting the template images from the CBCT images, residual images with image details and ring artifacts are generated. As the ring artifact manifests as a stripe artifact in a polar coordinate system, the artifact image can be extracted by mean value from the residual image; the image details are generated by subtracting the artifact image from the residual image …