Authors
Giovanna Ferraioli, Laura Maiocchi, Giovanni Savietto, Carmine Tinelli, Mara Nichetti, Mariangela Rondanelli, Fabrizio Calliada, Lorenzo Preda, Carlo Filice
Publication date
2021/7
Journal
Journal of Ultrasound in Medicine
Volume
40
Issue
7
Pages
1325-1332
Publisher
John Wiley & Sons, Inc.
Description
Objectives
The main aim was to assess the performance and cutoff value for the detection of liver steatosis (grade S > 0) with the Attenuation Imaging–Penetration (ATI‐Pen) algorithm available on the Aplio i‐series ultrasound systems (Canon Medical Systems, Otawara, Japan). The magnetic resonance imaging–derived proton density fat fraction (MRI‐PDFF) was used as the reference standard. Secondary aims were to compare the results to those obtained with the previous ATI algorithm (Attenuation Imaging–General [ATI‐Gen]) and with the controlled attenuation parameter (CAP) and to generate a regression equation between ATI‐Pen and ATI‐Gen values.
Methods
Consecutive adult patients potentially at risk of liver steatosis were prospectively enrolled. Each patient underwent ultrasound quantification of liver steatosis with ATI‐Pen and ATI‐Gen and a CAP assessment with the FibroScan system …
Total citations
202120222023202412313316
Scholar articles
G Ferraioli, L Maiocchi, G Savietto, C Tinelli, M Nichetti… - Journal of Ultrasound in Medicine, 2021