Authors
Cristian C Constantinescu, Jogeshwar Mukherjee
Publication date
2009/4/21
Journal
Physics in Medicine & Biology
Volume
54
Issue
9
Pages
2885
Publisher
IOP Publishing
Description
We evaluated the performance of an Inveon preclinical PET scanner (Siemens Medical Solutions), the latest MicroPET system. Spatial resolution was measured with a glass capillary tube (0.26 mm inside diameter, 0.29 mm wall thickness) filled with 18 F solution. Transaxial and axial resolutions were measured with the source placed parallel and perpendicular to the axis of the scanner. The sensitivity of the scanner was measured with a 22 Na point source, placed on the animal bed and positioned at different offsets from the center of the field of view (FOV), as well as at different energy and coincidence windows. The noise equivalent count rates (NECR) and the system scatter fraction were measured using rat-like (Φ= 60, L= 150 mm) and mouse-like (Φ= 25 mm, L= 70 mm) cylindrical phantoms. Line sources filled with high activity 18 F (> 250 MBq) were inserted parallel to the axes of the phantoms (13.5 and 10 …
Total citations
200920102011201220132014201520162017201820192020202120222023202411416182021122111131198723
Scholar articles
CC Constantinescu, J Mukherjee - Physics in Medicine & Biology, 2009