Nikolaos Melas, Theodosios Perdikides, Athanasios Saratzis, Nikolaos Saratzis, Dimitrios Kiskinis, David H Deaton
Publication date
2012/6/1
Journal
Journal of vascular surgery
Volume
55
Issue
6
Pages
1726-1733
Publisher
Mosby
Description
OBJECTIVE
This study evaluated the contribution of Aptus EndoStaples (Aptus Endosystems, Sunnyvale, Calif) in the proximal fixation of eight endografts used in the endovascular repair of abdominal aortic aneurysms (EVAR).
METHODS
Nine human cadaveric aortas were exposed, left in situ, and transected to serve as fixation zones. The Zenith (Cook, Bloomington, Ind), Anaconda (Vascutek, Inchinnan, Scotland, UK), Endurant (Medtronic, Minneapolis, Minn), Excluder (W. L. Gore and Associates, Flagstaff, Ariz), Aptus (Aptus Endosystems), Aorfix (Lombard Medical, Didcot, UK), Talent (Medtronic), and AneuRx (Medtronic) stent grafts were proximally deployed and caudal displacement force (DF) was applied via a force gauge, recording the DF required to dislocate each device ≥20 mm from the infrarenal neck. Measurements were repeated after four and six EndoStaples were applied at the proximal fixation …