Authors
Bathiya Prashan Senevirathna, Sheung Lu, Marc P Dandin, John Basile, Elisabeth Smela, Pamela A Abshire
Publication date
2018/6
Journal
IEEE transactions on biomedical circuits and systems
Volume
12
Issue
3
Pages
510-520
Publisher
IEEE
Description
We describe a capacitance sensor array that has been incorporated into a lab-on-CMOS system for applications in monitoring cell viability. This paper presents analytical models, calibration results, and measured experimental results of the biosensor. The sensor has been characterized and exhibits a sensitivity of 590 kHz/fF. We report results from benchtop tests and in vitro experiments demonstrating on-chip tracking of cell adhesion as well as monitoring of cell viability. Human ovarian cancer cells were cultured on chip, and measured capacitance responses were validated by comparison with images from photomicrographs of the chip surface. Analysis was performed to quantify cell proliferation and adhesion, and responses to live cells were estimated to be 100 aF/cell.
Total citations
2018201920202021202220232024456131174
Scholar articles
BP Senevirathna, S Lu, MP Dandin, J Basile, E Smela… - IEEE transactions on biomedical circuits and systems, 2018