Authors
Fahimeh Kamarei, Fabrice Gritti, Georges Guiochon, John Burchell
Publication date
2014/2/14
Journal
Journal of Chromatography A
Volume
1329
Pages
71-77
Publisher
Elsevier
Description
The implementation of the traditional FA method is difficult with classical supercritical fluid chromatography (SFC) instruments. The instrument mixer and other sources of extra-column volumes are large and significantly broaden the fronts of injected plugs, which diminishes the precision and accuracy of the FA method. An SFC instrument was modified to permit more accurate determinations of adsorption isotherm data. The sample, the modifier, and CO2 are separately pumped via small volume connection tubes into a small volume mixer (250 μL), where they are mixed into a homogeneous fluid fed to the column. The extra-column volumes and the column hold-up volume were accurately measured at each back pressure from the retention times of tracers. This modified instrument was used to measure the adsorption isotherm of S-naproxen by frontal analysis (FA) on a (R, R)-Whelk-O1 column, using a mixture of …